Softa (Eastofa) ELSD (Evaporative Light Scattering Detector) Test Report for Analysis of Lactose

Softa (ESF) ELSD ( Evaporative Light Scattering Detector)

Analysis of lactose test report (琛航科技)

First, the experimental conditions

Cometro High Performance Liquid Chromatography System;

Cometro (United States Connaught) 6000 LDI Pump ;

Softa ( Eastofa , USA) ELSD (Evaporative Light Scattering Detector), DT-70 °C, SC-30 °C;

Scienhome (Tianjin Airlines) Kromasil NH 2   Column

Mobile phase: acetonitrile - water = 70-30 ;

Flow rate: 1.0 ml/min ;

The concentration of lactose standard was 0.5mg/mL , 1mg/mL , 1.5mg/mL ;

Whey powder sample configuration: Take 0.1g powder into a 100mL volumetric flask   , dilute to the mark with water.

The injection volume is 20ul .

Second, the experimental results

Sample one:

Detector 1 Results

Retention Time

Area

Theoretical plates (USP)

Resolution (USP)

2.698

104649

1083

0.00000

16.797

3671699

5914

23.46733

Sample 2:

Detector 1 Results

Retention Time

Area

Theoretical plates (USP)

Resolution (USP)

2.695

149923

792

0.00000

16.871

3700293

5739

22.25627

Sample three:

Detector 1 Results

Retention Time

Area

Theoretical plates (USP)

Resolution (USP)

2.688

133156

845

0.00000

16.821

3 733804

5535

22.18048

Lactose standard (three injections at different concentrations, additional re-injection at a concentration of 1 mg/ml for 2 replicates):

Concentration and peak area relationship:

concentration

Peak area

Unknown (sample one)

3671699

Unknown (sample 2)

3700293

Unknown (sample 3)

3733804

Standard 0.5mg/mL

2288817

Standard 1.0mg/mL

4865288/4844636/4797327 RSD=0.72%

Standard 1.5mg/mL

7750119

With lg concentration as the abscissa and lg peak area as the ordinate

Bring the concentration of the sample into the three sample concentrations that can be calculated by the equation, as well as the mass percentage relative to the powder:

Concentration mg/mL

Mass percentage

Sample one

0.769619

76.96%

Sample two

0.775031

77.50%

Sample three

0.782605

78.26%

Third, the results of the discussion

It can be seen from the figure that the peak shape of the sample is good and the column efficiency is high. Softa (ESF) ELSD (evaporative light scattering detector) has good reproducibility (RSD=0.72%) , and the linear relationship is good ( R 2 =0.999) .