Softa (ESF) ELSD ( Evaporative Light Scattering Detector)
Analysis of lactose test report (ç›èˆªç§‘技)
First, the experimental conditions
Cometro High Performance Liquid Chromatography System;
Cometro (United States Connaught) 6000 LDI Pump ;
Softa ( Eastofa , USA) ELSD (Evaporative Light Scattering Detector), DT-70 °C, SC-30 °C;
Scienhome (Tianjin Airlines) Kromasil NH 2 Â Column
Mobile phase: acetonitrile - water = 70-30 ;
Flow rate: 1.0 ml/min ;
The concentration of lactose standard was 0.5mg/mL , 1mg/mL , 1.5mg/mL ;
Whey powder sample configuration: Take 0.1g powder into a 100mL volumetric flask  , dilute to the mark with water.
The injection volume is 20ul .
Second, the experimental results
Sample one:
Detector 1 Results | |||
Retention Time | Area | Theoretical plates (USP) | Resolution (USP) |
2.698 | 104649 | 1083 | 0.00000 |
16.797 | 3671699 | 5914 | 23.46733 |
Sample 2:
Detector 1 Results | |||
Retention Time | Area | Theoretical plates (USP) | Resolution (USP) |
2.695 | 149923 | 792 | 0.00000 |
16.871 | 3700293 | 5739 | 22.25627 |
Sample three:
Detector 1 Results | |||
Retention Time | Area | Theoretical plates (USP) | Resolution (USP) |
2.688 | 133156 | 845 | 0.00000 |
16.821 | 3 733804 | 5535 | 22.18048 |
Lactose standard (three injections at different concentrations, additional re-injection at a concentration of 1 mg/ml for 2 replicates):
Concentration and peak area relationship:
concentration | Peak area |
Unknown (sample one) | 3671699 |
Unknown (sample 2) | 3700293 |
Unknown (sample 3) | 3733804 |
Standard 0.5mg/mL | 2288817 |
Standard 1.0mg/mL | 4865288/4844636/4797327 RSD=0.72% |
Standard 1.5mg/mL | 7750119 |
With lg concentration as the abscissa and lg peak area as the ordinate
Bring the concentration of the sample into the three sample concentrations that can be calculated by the equation, as well as the mass percentage relative to the powder:
Concentration mg/mL | Mass percentage | |
Sample one | 0.769619 | 76.96% |
Sample two | 0.775031 | 77.50% |
Sample three | 0.782605 | 78.26% |
Third, the results of the discussion
It can be seen from the figure that the peak shape of the sample is good and the column efficiency is high. Softa (ESF) ELSD (evaporative light scattering detector) has good reproducibility (RSD=0.72%) , and the linear relationship is good ( R 2 =0.999) .